© Physik Instrumente (PI) GmbH & Co. KG Auf der Römerstr. 1 ⋅ 76228 Karlsruhe, Germany Tel. +49 721 4846-0 ⋅ Fax: +49 721 4846-299 info@pi
Introduction ! CAUTION Your system will be fully calibrated before being shipped. If you have informed PI about your application, calibration is d
Introduction 1.4 Model Survey The following E-517 models are available: E-517.i3 Interface / Display Module, 24 Bit D/A, TCP/IP, USB, RS-232, 3
Introduction MDR connector for Digital In/Out socket, consisting of: 4347 MDR connector, 14-pol, RM 1.27 4348 MDR cover 14-pol E-517.CD Distr
Introduction 1.7 Software Description Operating system details: "Windows" stands for 2000, XP and Vista "Linux" stands
Introduction Software ToolSupported Operating SystemShort Description Recommended forPIMikroMove™ Windows PIMikroMove™ permits you to start your m
First Steps 2 First Steps 2.1 Installing the Software on the Host PC Windows operating systems: 1 Insert the E-517 CD in your host PC. 2 If th
First Steps program. Proceed as follows to create a parameter file: 1 Power on the system. Find details in a dedicated section (p. 31). 2 Start
First Steps to copy the values of protected parameters from the Edit Mask column to the controller, enter the password "advanced" to swit
First Steps 3 Start PIMikroMove™ on the host PC. It opens with the Start up controller window. 4 Establish a connection to the E-517 from PIMikr
First Steps 5 Give the E-517 interface / display module control over the output voltage channels of the piezo control electronics by switching the
First Steps Find details in "Modes of Operation" (p. 23). 6 Optionally: Configure the PIMikroMove™ main window: It is recommended to s
First Steps 7 Make some test moves with the individual axes using the controls on the Axes tab card: The first moves should be made in open-loop
First Steps figure below). In closed-loop operation, target positions for the axes are commanded. Enter the target position in the Target field and
Operation 3 Operation NOTES All commands mentioned in the subsections below are PI General Command Set (GCS) commands in the E-517 syntax, provide
Operation 3.1 Front Panel Elements 3.1.1 Front Panel Overview Figure 1: E-517.i3 interface / display module, 3 channels. Single-channel E-517.i1
Operation 3.1.2 Display Screens The individual display screens can be accessed by the trackball (p. 22) on the E-517 front panel. You can check th
Operation CHANNEL SETTING, where you can command motion (in ONLINE mode) or the check current position and voltage (in OFFLINE mode) for individ
Operation In OFFLINE mode, you can check the current voltage and position values on the channel-specific screens. 3.2 Modes of Operation Contr
Operation See "Control Value Generation" in the E-517 User Manual for details. Switching between ONLINE and OFFLINE mode can be done by:
Operation operation omits the servo loop, and the control input directly controls the output voltage of the piezo channel. The slew rate can be inf
Physik Instrumente (PI) GmbH & Co. KG is the owner of the following company names and trademarks: PI®,
Operation On power-on or reboot (with the RBT command), the E-517 copies information from non-volatile memory to volatile memory. The control mode
Operation While move commands and trackball settings can be used in parallel to command axis motion, they are not accepted when the wave generator
Operation Hardware: A piezo stage and an E-500 chassis with one E-505 amplifier module and E-517 (single-channel system). Command String to Send R
Operation Command String to Send Response Action Performed SVA? A A=+0085.0000 Query the commanded output voltage. It resulted from the last vali
Operation Command String to Send Response Action Performed ONL 1 1 2 1 3 1 Switch all piezo channels (identifiers 1, 2, 3) to ONLINE mode so that
Operation 3.5 Data Recording 3.5.1 How to Use the Data Recorder The E-517 includes a real-time data recorder. It is able to record several input
Operation The data recorder sampling period can be read with the RTR? command. The answer gives the value of the Data Recorder Table Rate parameter
Operation Send: SPA 1 0x05000000 0 to deactivate the digital filter. Close the Command Entry window. 4 Open the Data Recorder window via the E-517
Operation 8 To analyze the recorded data and display it as a Bode frequency response diagram, click on the FFT button in the icon bar above the gr
Operation 3.6 Using Trigger Input and Output It is possible to trigger external devices and to send start/stop signals to the wave generators wit
About This DocumentUsers of This Manual This Quick Guide is designed to help the reader to put the E-517 Digital Piezo Controller Operation Module int
Operation 3.7 How to Work with the Wave Generator Each axis can be controlled by a "wave generator" which outputs user-specified pattern
Operation 3 Stop the wave generator output with WGO or #24 or STP. The simple example below shows how to start periodic motion of the first axis (
Operation 3.8.1 Defining Macros To define a macro command sequence, first activate macro recording mode with the command MAC BEG <macroname>
Operation Example 2: Macro1 now is splitted in two separate macros (macro2 and macro3) because it in not necessary to initialize the E-517 every ti
Operation Macro execution can be stopped from the command line with #24. The STP command stops macro execution only when it is part of the macro.
Operation 3.9 Select Command Set Version Users who have written software for the E-516 interface / display module which is the predecessor of the
Operation 3.10 Updates 3.10.1 Software Updates Updated releases of software and manuals are available for download at www.pi.ws. While the manua
Operation All firmware updates can be made by running the Firmware Update Wizard on the host computer, using the same communication interfaces as f
Operation Figure 3: Select controller to be connected to Firmware Update Wizard www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 44
Communication 4 Communication 4.1 Interfaces Available The E-517 can be controlled from a host computer (not included) with ASCII commands sent v
Contents 1 Introduction 3 1.1 How to Use the E-517 Documentation ...4 1.2 Prescribed Use...
Communication NOTES All commands mentioned in the subsections below are PI General Command Set (GCS) commands in the E-517 syntax, provided by the
Communication GPADR: gives the device address to be used for GPIB (IEEE 488) communication, default is 4 also accessible as GPIB Address paramet
Communication 4.3 TCP/IP Connection The TCP/IP connection is available on the E-517 front panel of the controller, via the RJ 45 socket with the n
Communication First, make sure that the proper startup behaviour is set (see Note above). Then, if a network with DHCP server is available, connect
Communication 4.3.2 PC with Ethernet Card The IP address and IP mask settings of PC and E-517 must be compatible with each other in the following
Communication 3 In the Properties window of the LAN connection, click on Internet Protocol (TCP/IP). 4 In the Internet Protocol (TCP/IP) Properties
Communication Adapt IP address and Subnet mask; do not change the Default gateway setting Press OK Figure 5: Internet Protocol (TCP/IP) Pr
Communication Do not use "255", and do not change the port setting (must always be 50000). 3 Close the connection. 4 Connect the E-517
Communication Figure 6: RS-232 configuration of the host PC side in PIMikroMove™ If you have established a connection and want to change the curr
Communication 4.5 USB Connection The USB interface is available on the front panel of the E-517 via the type B USB socket. Use the included USB ca
3.10.1 Software Updates ... 42 3.10.2 Firmware Updates ...
Communication 4.6 GPIB (IEEE 488) Parallel Connection The parallel communications bus is accessed via the "GPIB/IEEE488" connector on
Communication If you have established a connection and want to change the currently active address of the E-517, proceed as follows: 1 Use the I
E-517 GCS Commands, Alphabetical Survey 5 E-517 GCS Commands, Alphabetical Survey #5 Request Motion Status #6 Query If Position Has Changed Sin
E-517 GCS Commands, Alphabetical Survey DRC Set Data Recorder Configuration DRC? Get Data Recorder Configuration DRR? Get Recorded Data Values
E-517 GCS Commands, Alphabetical Survey ONT? Get On Target State OVF? Get Overflow State PLM Set High Position Soft Limit PLM? Get High Posi
E-517 GCS Commands, Alphabetical Survey TIO? Tell Digital I/O Lines TMN? Get Minimum Commandable Position TMX? Get Maximum Commandable Positio
E-517 GCS Commands, Alphabetical Survey WGC Set Number Of Wave Generator Cycles WGC? Get Number Of Wave Generator Cycles WGO Set Wave Generato
TTroubleshooting 6 Troubleshooting Communication with controller does not work Communication cable is wrong or defective ⇒ Check cable. Does it wo
TTroubleshooting ⇒ The first time you connect over the USB interface, be sure you are logged on the PC as a user having administrator rights. After
TTroubleshooting be identical or the parameters of the GEMAC interpolation circuit must be adapted, see "GEMAC Parameter Adjustment" for
Introduction 1 Introduction Low-Noise 24-bit D/A Converter Sample Rate 25 kHz TCP/IP, USB, IEEE 488 and RS-232 Interfaces 6-Digit Disp
TTroubleshooting Incorrect control mode of the piezo channel ⇒ Check the current control mode using the ONL? command or on the main screen of the E
TTroubleshooting When using the wave generator, it is recommended to reduce the frequency and/or the amplitude and/or the output duration to avoid
Customer Service 7 Customer Service Call your PI representative or write to [email protected]; please have the following information about your system re
Old Equipment Disposal 8 Old Equipment Disposal In accordance with EU directive 2002 / 96 / EC (WEEE), as of 13 August 2005, electrical and electr
Technical Data 9 Technical Data 9.1 Specifications E-517.i1 E-517.i3 Function Digital operation module Digital operation module Channels 1
Technical Data 9.2 Pin Assignments 9.2.1 Digital In/Out Socket Connector type: MDR14 Pin Signal Function 1 GND GND 2 nc not connected 3 out
Technical Data 9.2.3 Main Connector Connector type: 32-pin connector, DIN 41612 Row Pin a Pin c 1 AGND AGND 2 IN: ch1 OUT: ch1 3 IN: ch2 OU
Introduction 1.1 How to Use the E-517 Documentation This Quick Guide enables you to put the E-517 into operation for the first time and gives you
Introduction 1.2 Prescribed Use The E-517 Digital Piezo Controller Operation Module is designed for installation in E-500/E-501, E-471 and E-481 m