E-Lins E481 Series Specifikace Strana 71

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 76
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 70
TTroubleshooting
When using the wave generator, it is recommended to reduce the
frequency and/or the amplitude and/or the output duration to avoid
overheating. See "Wave Generator" in the E-517 User Manual for more
information.
Unsatisfactory system performance
The sensor values are not reliable, and the whole system is
instable.
Only thermally stable systems can have the best performance. For a
thermally stable system, power on the E-517 at least one hour before you
start working with it.
Custom software accessing PI drivers does not run.
Wrong combination of driver routines/VIs
Check if system runs with Terminal program. If yes read the software
manual and compare sample code from the E-517 CD to check the
necessary driver routines.
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 67
Zobrazit stránku 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76

Komentáře k této Příručce

Žádné komentáře